| Model Number |
106762 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient had an incision and drainage procedure.The primary system controller was exchanged in the operating room.Log files were sent for both primary and backup controller.While in the operating room the primary controller at the time displayed the broken line (---) for its flow reading.As a trouble shooting measure by the perfusion team, the primary controller was exchanged for the patient's backup controller.Once the new controller was connected to patient it also displayed (---) in the flow reading.The physician thought it could be an issue with the driveline.No alarms were noted at the time.It was communicated by the perfusionist there was an alarm noted in the historical view on 06jan2026, indicating for primary controller to be exchanged, which never happened.The log file captured several low-speed advisories and pump stops on 06jan2026 while the patient was connected to the power module.These alarms may have been the result of a short within the driveline.It was recommended to keep the vad connected to battery power or an ungrounded cable.It was noted that this patient had a driveline repair in jun2022 for cosmetic damage ((b)(4)) so another repair would not be possible.The driveline was disconnected at 1:36 pm on 13jan2026 for a controller exchange.The log file for the backup controller captured low flow events on 13jan2026.There did not appear to be any type of mechanical circulatory support (mcs) equipment issues causing these events.The low flow events seemed to be a patient related issue and not an equipment related issue.There were no other unusual events recorded in the log file event history.The mcs equipment was operating as intended.The patient was placed on veno-arterial extracorporeal membrane oxygenation (va ecmo).The physician requested a picture of the driveline to assess if there was enough room for a percutaneous lead repair.The driveline repair was scheduled for 16jan2026 due to suspected short to shield.X-ray images were taken.The repair took place.It was noted that there was approximately 5" between new exit site and existing repair site.The outer layer and bionate were removed between 3" and 5" from exit site, right up to existing repair.The green, brown, and orange wires were spliced, and the drivelines were swapped without issue.Splicing was continued on yellow, black, and red wires.The area was closed up per procedure, and the patient was provided with care instructions.
|
| |
|
Additional Manufacturer Narrative
|
|
Device serial number and lot number were not provided, and expiration date and manufacture date (h4) cannot be determined.The primary udi number in d4 is reflective of all available information.H10: no further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is complete.
|
| |
|
Event or Problem Description
|
|
More recent log files were submitted, and they captured multiple low speed and pump stop alarms on (b)(6) 2026 while the patient was connected to the power module.This indicated that the repair on (b)(6) 2026 did not remove the damaged area and is likely internal.The patient would receive an ungrounded cable to prevent alarms.The patient was on battery power and was doing okay.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation conclusion: the heartmate ii system controller was evaluated following the reported event of a replace system controller alarm; however, it was determined that the system controller was unrelated to the cause of the reported event.The submitted log file contained approximately 7 days of data (b)(6) 2026 per the timestamp).The log file captured a replace system controller alarm on (b)(6) 2026 from 16:08:34 to 16:09:36 that was associated with a backup mcu fault.This alarm is consistent with the damage to the driveline and is addressed further under the pump investigation.The reported event could not be correlated to an issue with the system controller.The serial number of the heartmate ii system controller was not reported with the event.The current revision of the instructions for use (ifu) can be found on the eifu page of the abbott website.Heartmate ii instructions for use section 7 ¿ ¿alarms and troubleshooting¿ and heartmate ii patient handbook section 5 ¿ ¿alarms and troubleshooting¿ cover all alarms (visual and auditory), including replace system controller alarms, and the actions to take if the issue does not resolve.The patient handbook and the ifu caution the user to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
|
| |
|
Search Alerts/Recalls
|