• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ULTRA FAST-FIX AB ASSEMBLY - CURVED; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ULTRA FAST-FIX AB ASSEMBLY - CURVED; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE Back to Search Results
Catalog Number 72201494
Medical Device Problem Code Premature Activation (1484)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 12/29/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that, during a meniscus repair, the t2 implant of one (1) ultra fast-fix curved came off the inserter when trying to advance it to the deployment position, the t1 implant was left secured inside the patient.The procedure was resumed, with a non-significant delay, using an equivalent s+n backup.No further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRA FAST-FIX AB ASSEMBLY - CURVED
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key24179678
Report Number1219602-2026-00238
Device Sequence Number9578647
Product Code GAT
UDI-Device Identifier03596010597229
UDI-Public3596010597229
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K151105
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72201494
Device Lot Number2164145
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2026
Initial Report FDA Received Date01/26/2026
Date Device Manufactured09/02/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-