| Catalog Number |
1115000000E |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Events occurred between october 1-december 31, 2025.This report summarizes 2 malfunction event(s), where it was reported the device experienced steer mechanism is difficult to engage/disengage.No adverse consequence or clinically relevant delay in treatment was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.1 device(s) was/were functionally/visually inspected in the field.The device(s) was/were repaired and returned to use.1 device was not evaluated, as a probable cause for the issue was identified during communication between the customer and stryker and no further assistance was requested by the customer.Evaluation results findings (components/results) 2 devices: rod/shaft / mechanical problem identified.There was no remedial action taken.This device is not labeled for single use.
|
| |
|