| Catalog Number |
FL19H |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Events occurred between october 1-december 31, 2025.This report summarizes 2 malfunction event(s), where it was reported the device experienced inaccurate information on scales.No adverse consequence or clinically relevant delay in treatment was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.1 device(s) was/were functionally/visually inspected in the field.The device(s) was/were repaired and returned to use.Evaluation results findings (components/results) 1 device: sensor / failure to calibrate or used out of calibration.1 device was not evaluated and no cause was determined, as the customer did not make the device accessible for testing.Evaluation results findings (components/results) 1 device: appropriate term/code not available / no findings available.There was no remedial action taken.This device is not labeled for single use.
|
| |
|
Search Alerts/Recalls
|