Catalog Number FL19H
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Events occurred between october 1-december 31, 2025.This report summarizes 5 malfunction event(s), where it was reported the device experienced end/siderail cannot latch upright, mid-position, or stuck at lowest height.No adverse consequence or clinically relevant delay in treatment was reported.
 
Additional Manufacturer Narrative
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.5 device(s) was/were functionally/visually inspected in the field.The device(s) was/were repaired and returned to use.Evaluation results findings (components/results) 2 devices: side rail / dust or dirt problem identified 1 device: cable; mechanical/structural / mechanical problem identified 1 device: side rail / wear problem 1 device: side rail / mechanical problem identified.There was no remedial action taken.This device is not labeled for single use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUB PEDIATRIC CRIB- HYD BASE
Common Device Name
BED, PEDIATRIC OPEN HOSPITAL
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
melissa simon
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key24183704
Report Number0001831750-2026-99084
Device Sequence Number9557624
Product Code FMS
UDI-Device Identifier07613327169348
UDI-Public7613327169348
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Exemption NumberVMSR
Number of Events Summarized5
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberFL19H
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/01/2026
Initial Report FDA Received Date01/26/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-