| Catalog Number |
HARH45 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 1/26/2026.D4 batch #: unknown.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown procedure the white tissue pad was completely detached.The fallen piece was retrieved by forceps and was disposed.Changed to another device to complete surgery.There was no patient consequence reported.No additional information can be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 3/27/2026.D4 batch #: a97f48.Investigation summary- the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the device was returned with the tissue pad detached and not returned.The blade tip was damaged, however, the blade was not cracked.The device was connected to the energy system and the device did activate during functional testing.The device was disassembled to verify the condition of the internal components and no anomalies were found.Based on the condition of the tissue pad, a probable cause of this damage is that the clamp of the device may have been closed and the instrument activated without tissue present.Care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Keep the clamp arm open when back cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad.The resulting damage contributes to the removal of the pad from the clamp arm.The cleaning of the pad, not in accordance with the ifu, can also result in removal of the pad during use.Probable causes of the blade damage are applying pressure between the instrument blade and tissue pad without having tissue between them, subsequent activations would completely wear and melt the tissue pad until the blade tip makes contact with the clamp arm.Avoid activating the blade without tissue between the blade and tissue pad to prevent this type of damage.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch a97f48, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|