| Catalog Number |
1460000000 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Events occurred between october 1-december 31, 2025.This report summarizes 4 malfunction event(s), where it was reported the device experienced reduced/inadequate brake force.No adverse consequence or clinically relevant delay in treatment was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.2 device(s) was/were functionally/visually inspected in the field.The device(s) was/were repaired and returned to use.Evaluation results findings (components/results): 1 device: actuator / wear problem.1 device: fastener / mechanical problem identified.1 device was not evaluated and no cause was determined, as the customer did not make the device accessible for testing.Evaluation results findings (components/results): 1 device: appropriate term/code not available / no findings available.1 device is pending evaluation.Evaluation results findings (components/results): 1 device: insufficient information / results pending completion of investigation.There was no remedial action taken.This device is not labeled for single use.
|
| |
|
Event or Problem Description
|
|
No new information.
|
| |
|
Additional Manufacturer Narrative
|
|
The device that was pending evaluation was evaluated in the field and the issue was confirmed.The device was repaired on site and returned to service.Evaluation results findings (components/results).1 device: fastener/device migration.
|
| |
|
Search Alerts/Recalls
|