| Catalog Number |
367988 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There were multiple 510k numbers reported to be involved.The information is as follows: g.4.Pma / 510(k)#: k230855.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® sst¿, an unspecified number of samples exhibit elevated potassium results.There was no health impact or consequences reported.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® sst¿, an unspecified number of samples exhibit elevated potassium results.There was no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Bd did not receive and samples or photos for investigation.Factors that may contribute to erroneous results were evaluated through a clinical study to verify the design of the device met it¿s intended use.Bd was unable to duplicate the customer¿s indicated failure mode (erroneous results-potassium) because the defect was not evident in the testing of the complaint lot samples.Replicates of both customer retain and control samples were acceptable in terms of both precision and accuracy.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode erroneous results-potassium.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Search Alerts/Recalls
|