STAAR SURGICAL COMPANY EVO/EVO+ VISIAN IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM5_13.2(-13.5/2.0)
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
B5 - add info received: the problem is resolved - perfect patient condition at the post op exam.Claim # (b)(4).
 
Additional Manufacturer Narrative
B5 - the reporter indicated that following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault, elevated iop and significant reduction of iridocorneal angles.Due to excessive vault, elevated iop and significant reduction of iridocorneal angles the lens was exchanged for a same model, power but shorter length lens.Cause of the event is unknown.G3 - date received by manufacturer: 20-jan-2026 correct to 06-jan-2026 h3 - device evaluation: the lens was returned dry in a vial with residue/debris on the lens.Visual inspection found the lens optic deformed and haptic deformed and stretched out.Dimensional inspection found the lens to be within specifications.Claim #(b)(4).
 
Additional Manufacturer Narrative
H6 - 4581: significant reduction of ica.H6 - work order search: no additional similar complaint type events within associated lots were found.Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim # (b)(4).
 
Event or Problem Description
The reporter indicated that following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault, elevated iop & tight angle.Due to excessive vault, elevated iop and tight angle the lens was exchanged for a same model but shorter length lens.Cause of the event is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVO/EVO+ VISIAN IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
MDR Report Key24185397
Report Number2023826-2026-00119
Device Sequence Number9649900
Product Code QCB
UDI-Device Identifier00840311312085
UDI-Public(01)00840311312085(11)250531(17)280430
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberVTICM5_13.2(-13.5/2.0)
Device Catalogue NumberVTICM5_13.2
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/20/2026
Supplement Date Received by Manufacturer03/17/2026
05/04/2026
Initial Report FDA Received Date01/26/2026
Supplement Report FDA Received Date03/30/2026
05/07/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/31/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL-SFC-45, LOT#-UNK; FOAM TIP PLUNGER-MODEL-UNK, LOT#-UNK; INJECTOR MODEL-UNK,LOT#-UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-