MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BARDEX® PEDIATRIC ALL-SILICONE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BARDEX® PEDIATRIC ALL-SILICONE FOLEY CATHETER Back to Search Results
Catalog Number 165808
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 01/08/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the balloon on the foley catheter would not deflate on two different occasions.The first time the doctor was able to deflate that using mineral oil to dissolve that (lot- ngkv0323) and then the second time the doctor office was closed due to the holidays which caused them to have to go to the emergency room.This was a pediatric patient.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the balloon on the foley catheter would not deflate on two different occasions.The first time the doctor was able to deflate that using mineral oil to dissolve that (lot ngkv0323) and then the second time the doctor office was closed due to the holidays which caused them to had to go to the emergency room.This was a pediatric patient.
 
Additional Manufacturer Narrative
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.The reported event is addressed within the labeling as it states: warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage the catheter and may cause balloon to burst.Precautions: do not use device if package is opened or damaged.Do not aspirate urine through drainage funnel wall.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Instructions for use: visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Removal: wash hands.Gently insert a luer slip tip syringe into the catheter valve.Never use more force than is required to make the syringe "stick" in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe.If you notice slow or no deflation, reseat the syringe gently.Use only gentle aspiration to encourage deflation, if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If balloon does not deflate, contact trained professional for assistance, as directed by hospital protocol.After balloon is fully deflated, remove catheter, and dispose of in accordance with hospital policy.A dhr could not be performed since no lot number was provided.No additional actions can be taken at this time.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® PEDIATRIC ALL-SILICONE FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key24185878
Report Number1018233-2026-00243
Device Sequence Number9575137
Product Code KOD
UDI-Device Identifier00801741029714
UDI-Public(01)00801741029714
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K821745
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number165808
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2026
Supplement Date Received by Manufacturer04/16/2026
Initial Report FDA Received Date01/26/2026
Supplement Report FDA Received Date04/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/28/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-