| Model Number |
OPT316 |
| Medical Device Problem Codes |
Gas/Air Leak (2946); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Fisher & paykel healthcare (f&p) is currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.
|
| |
|
Event or Problem Description
|
|
A distributor in china reported on behalf of a healthcare facility that the opt316 optiflow junior nasal cannula was leaking air during use.It was further reported that the tubing of the nasal cannula was damaged at the connector (swivel grip) end.There was no reported patient consequence.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Method: the subject device, opt316 optiflow junior nasal cannula was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the information and photography provided by the distributor and our knowledge of the product.Results: evaluation of the provided photography confirmed that both tubes were stretched and torn at the connector (swivel grip) end.Additionally, it was observed that part of the connector (swivel grip) was missing.The distributor reported that the subject device was connected to an rt329 single limb infant breathing circuit kit.Conclusion: based on the evaluation of photography provided, and our knowledge of the product, it is most likely that the reported event resulted from the cannula being subjected to excessive force.The missing part in the connector may have been removed to enable connection to a breathing circuit other than the one recommended in the user instructions.As part of the manufacturing process, all optiflow junior cannulas are 100% leak and occlusion tested after final assembly and any cannula that fails specifications is discarded.In addition, representative samples from each batch are functionally tested to assess retention strength between the assembled components.If there are any failures the entire batch is quarantined for further investigation.The user instructions which accompany the optiflow junior nasal cannula show in pictorial format the correct placement and fitting of the cannula, and provide the following warnings: - "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death." - "ensure that all connections are secure dung use.Check cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient." - "do not stretch or crush tube." the subject device is intended to be used with f&p rt330 optiflow¿ junior breathing circuit kit when used with mr850 heated humidifier.
|
| |
|
Event or Problem Description
|
|
A distributor in china reported on behalf of a healthcare facility that the opt316 optiflow junior nasal cannula was leaking air during use.It was further reported that the tubing of the nasal cannula was damaged at the connector (swivel grip) end.There was no reported patient consequence.
|
| |
|