| Model Number |
CCA0T0 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that during an intraocular lens (iol) implant procedure, it was noted that the plunger went passed the lens.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in h.3., h.6.And h.11.The product was not returned for analysis.The complainant indicates the use of healon as a viscoelastic, which is not qualified for use with associated model, this is not a qualified company product combination.The reported complaint was not observed as no sample was returned for analysis; however, based on the information provided by the customer, the most likely root cause is failure to follow instructions for use (ifu), as the surgeon states the use of non-qualified viscoelastic.The use of an unqualified combination may cause damage to the iol and potential complications during the implantation process.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
|
| |
|