| Catalog Number |
300900000000 |
| Medical Device Problem Code |
No Audible Alarm (1019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Events occurred between october 1-december 31, 2025.This report summarizes 3 malfunction event(s), where it was reported the device experienced clinician not alerted/aware of possible patient fall.No adverse consequence or clinically relevant delay in treatment was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.3 device(s) was/were functionally/visually inspected in the field.The device(s) was/were repaired and returned to use.Evaluation results findings (components/results).1 device: wiring harness / electrical/electronic component problem identified.1 device: switch/relay / mechanical problem identified.1 device: cable; electrical / electrical/electronic component problem identified.There was no remedial action taken.This device is not labeled for single use.
|
| |
|