|
Additional information: d9, g3, h2, h3, h6, h11.One ampere¿ rf ablation generator was received for evaluation.The product functioned as anticipated during testing with no abnormalities identified that would result in the reported complaint.Based on the information provided to abbott and the investigation performed, root cause of the field reported event could not be conclusively determined as the returned ampere generator functioned as anticipated, with no anomalies identified that would result in the reported complaint.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.
|