| Model Number |
7736 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Fracture (1260); High impedance (1291); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/29/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this right atrial (ra) lead exhibited oversensing of noisy signals and high out of range impedance measurements.The lead safety trigger switch had been triggered three months back, which cause the lead to switch from bipolar to unipolar.The ra lead was reported to have a fracture.Technical services (ts) was consulted and discussed stored data as well as available programming options.This device remains in service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this right atrial (ra) lead exhibited oversensing of noisy signals and high out of range impedance measurements.The lead safety trigger switch had been triggered three months back, which cause the lead to switch from bipolar to unipolar.The ra lead was reported to have a fracture.Technical services (ts) was consulted and discussed stored data as well as available programming options.This device remains in service.No adverse patient effects were reported.At a later date, this right atrial (ra) lead was capped and surgically abandoned.A new device was implanted.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of fracture was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.The device was not returned for analysis; therefore, a technical analysis could not be performed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|
| |
|
Event or Problem Description
|
|
It was reported that this right atrial (ra) lead exhibited oversensing of noisy signals and high out of range impedance measurements.The lead safety trigger switch had been triggered three months back, which cause the lead to switch from bipolar to unipolar.The ra lead was reported to have a fracture.Technical services (ts) was consulted and discussed stored data as well as available programming options.This device remains in service.No adverse patient effects were reported.At a later date, this right atrial (ra) lead was capped and surgically abandoned.A new device was implanted.No additional adverse patient effects were reported.
|
| |
|