Model Number 7736
Medical Device Problem Codes Signal Artifact/Noise (1036); Fracture (1260); High impedance (1291); Over-Sensing (1438)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/29/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this right atrial (ra) lead exhibited oversensing of noisy signals and high out of range impedance measurements.The lead safety trigger switch had been triggered three months back, which cause the lead to switch from bipolar to unipolar.The ra lead was reported to have a fracture.Technical services (ts) was consulted and discussed stored data as well as available programming options.This device remains in service.No adverse patient effects were reported.
 
Event or Problem Description
It was reported that this right atrial (ra) lead exhibited oversensing of noisy signals and high out of range impedance measurements.The lead safety trigger switch had been triggered three months back, which cause the lead to switch from bipolar to unipolar.The ra lead was reported to have a fracture.Technical services (ts) was consulted and discussed stored data as well as available programming options.This device remains in service.No adverse patient effects were reported.At a later date, this right atrial (ra) lead was capped and surgically abandoned.A new device was implanted.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of fracture was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.The device was not returned for analysis; therefore, a technical analysis could not be performed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
 
Event or Problem Description
It was reported that this right atrial (ra) lead exhibited oversensing of noisy signals and high out of range impedance measurements.The lead safety trigger switch had been triggered three months back, which cause the lead to switch from bipolar to unipolar.The ra lead was reported to have a fracture.Technical services (ts) was consulted and discussed stored data as well as available programming options.This device remains in service.No adverse patient effects were reported.At a later date, this right atrial (ra) lead was capped and surgically abandoned.A new device was implanted.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGEVITY? MRI
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key24189262
Report Number2124215-2026-04654
Device Sequence Number15470156
Product Code NVN
UDI-Device Identifier00802526523397
UDI-Public(01)00802526523397(17)200506
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P150012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/06/2020
Device Model Number7736
Device Catalogue Number7736
Device Lot Number682208
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2026
Supplement Date Received by Manufacturer01/27/2026
04/30/2026
Initial Report FDA Received Date01/27/2026
Supplement Report FDA Received Date02/02/2026
05/04/2026
Date Device Manufactured05/06/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age76 YR
Patient SexMale
-
-