• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. Q-SYTE; STOPCOCK, I.V. SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. Q-SYTE; STOPCOCK, I.V. SET Back to Search Results
Catalog Number 385102
Medical Device Problem Codes Break (1069); Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that bd q-syte ext set, luer-lok 6 in micro bore component damage and leaked the front end of a closed-system iv connector ruptured, causing medication leakage.Medical staff promptly replaced the connector upon discovery, preventing any harm to the patient.
 
Additional Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event or Problem Description
No additional information.
 
Additional Manufacturer Narrative
Investigation results: bd was not able to confirm the customer¿s indicated failure mode since no photo nor sample was provided to perform a proper investigation.An investigation on the actual device would be necessary to confirm the reported failure mode.Quality records have been consulted for tracking and trending purposes but issues like this are not detected which means pretty low occurrence.We will keep monitoring the manufacturing process and in case any emerging trend is detected, further actions will be taken if necessary.No related quality issues or deviations were found during the manufacture of the subassembly batches.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Q-SYTE
Common Device Name
STOPCOCK, I.V. SET
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
Manufacturer (Section G)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX  
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key24189506
Report Number9610847-2026-00031
Device Sequence Number8065081
Product Code FPA
UDI-Device Identifier00382903851027
UDI-Public(01)00382903851027(17)291031(10)5010180
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K013621
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number385102
Device Lot Number5010180
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2026
Supplement Date Received by Manufacturer02/24/2026
Initial Report FDA Received Date01/27/2026
Supplement Report FDA Received Date02/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/28/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-