| Model Number |
6946M62 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Failure to Capture (1081); High impedance (1291); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
Arrhythmia (1721)
|
| Date of Event |
01/05/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented to the emergency room (er) with palpitations.Upon review, it was noted that the right ventricular (rv) lead had triggered a lead integrity alert (lia) due to increased sensing integrity counter (sic) and high rate non-sustained episodes.In addition, the rv lead exhibited oversensing and noise.The rv lead also triggered an alert for undefined high pacing impedances.Loss of capture was also noted on the rv lead and the rv amplitude was set too high.The high voltage therapy was reprogrammed; ventricular fibrillation (vf) detection was programmed off.The rv lead remains in use.No further patient complications have been reported as a result of this event.
|
| |
|
Event or Problem Description
|
|
It was further reported that the rv lead was explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|