SYNTHES GMBH SMALL BATTERY DRIVE II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
532.110 |
| Medical Device Problem Codes |
Break (1069); Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the small battery drive device trigger moving parts did not move smoothly.It was reported in the service order that the device button which allowed the device to reverse, rotation was broken.It was unknown when the event occurred.There were no reports of delays to a surgical procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown but was known to have occurred in 2025.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11.Additional narrative: a visual and functional assessment was performed on the device.The following steps failed: general condition.Check for sticky triggers.Check function of device.Check power with power test bench.During the service evaluation the following failures were identified: functional: will not run - motor damaged.Functional: moving parts do not move smoothly - trigger.Visual: component damaged.Conclusion: failure reported is confirmed.
|
| |
|
Search Alerts/Recalls
|
|
|