| Catalog Number |
03.404.029S |
| Medical Device Problem Codes |
Break (1069); Entrapment of Device (1212); Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
11/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: d4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.
|
| |
|
Event or Problem Description
|
|
While using the ria 2 reamer irrigation system, the distal locking pins for the reamer head disconnected and couldn't be retrieved.Per the surgeon, they were backing out the reamer when the head disengaged, he attempted to reengage the reamer head and was unsuccessful, causing the locking pins to come off.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d4 (expiry date).Corrected: d4 (primary udi number).H3, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable device history review (dhr): manufacturing location: supplier ¿ (b)(4) / inspected and packaged by: monument.Release to warehouse date: 22-sep-2023.Expiration date: 31-aug-2033.Part number: 03.404.029s, 16.5mm reamer head for ria 2-sterile.Lot number: 7850p74 (sterile).Lot quantity: (b)(4).Nonconformance noted: n/a.A manufacturing record evaluation was performed for the finished device with batch number 7850p74.No nonconformities or manufacturing irregularities have identified related to the complaint condition.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|