| Model Number |
GIF-XP180N |
| Medical Device Problem Code |
Device Reprocessing Problem (1091)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/20/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: according to previous investigations, the use of products that contain silicone can cause deposits of white foreign material.Silicone is for example included in simethicone, a defoaming agent, lubricants, etc.Simethicone and petroleum/oil/silicone-based lubricants are non-water soluble and thus not recommended for use by olympus.The event can be prevented by following the instructions for use which states: nothing other than sterile water should be used for air/water feeding.No additives should be put into the sterile water.Non-sterile water may cause patient cross-contamination and/or infection.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the gastrointestinal videoscope had white foreign material in the air/water cylinder, air/water tube, nozzle, objective lens and plastic distal end cover.There was no patient involvement.
|
| |
|