| Model Number |
MT24078 |
| Medical Device Problem Code |
Low Audible Alarm (1016)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that an alert/notification settings issue occurred.No product or data was provided for evaluation.The allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that an alert/notification settings issue occurred.An external visual inspection was performed and passed.Receiver charge and boot was performed and passed.Data log analysis was performed and passed.Functional tests were performed and passed.Alarm/alert manual test was performed and passed.Speaker/vibrator resistance was performed and passed.An internal visual inspection was performed and passed.Performance data was reviewed.An alert/ notification settings issue was not found within the investigation window.The probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|