• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: UNITED ORTHOPEDIC CORPORATION TIBIAL TOWER NON-CEM #6-#7; TIBIAL GUIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNITED ORTHOPEDIC CORPORATION TIBIAL TOWER NON-CEM #6-#7; TIBIAL GUIDE Back to Search Results
Model Number 9404-2903
Medical Device Problem Codes Break (1069); Material Separation (1562); No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The problem may be caused by an inappropriate usage.While encountering to dense bone, the surgeon should use the tibial peg drill to make four pilot holes before positioning the tibial tower onto the tibial baseplate trial, otherwise the device breakage might happened.On the other hands, the device manufacturing manner of the spike part has been revised from onepiece machining to welding to further enhance the device strength.
 
Event or Problem Description
The surgeon removed the tibial tower component #6~#7 after applying force (punching) to the patient's knee.Upon inspection, the surgeon found that one of the spikes within the tibial tower was broken and had become embedded in patient's bone.The surgeon utilized a rongeur tool to carefully remove the broken spike fragments from patient's bone.After removing the broken spike, the surgeon proceeded to complete the case without further incidents.The tibial tower had already served its intended purpose prior to its removal, so the broken spike did not compromise the surgical outcome.There was a brief delay of 30 seconds which did not contribute to any impact on the patient or the overall success of the surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL TOWER NON-CEM #6-#7
Common Device Name
TIBIAL GUIDE
Manufacturer (Section D)
UNITED ORTHOPEDIC CORPORATION
no. 57, park ave. 2
science park
hsinchu city, 30075
TW  30075
Manufacturer (Section G)
UNITED ORTHOPEDIC CORPORATION
no. 57, park ave. 2
science park
hsinchu city, 30075
TW   30075
Manufacturer Contact
lois ho
no. 57, park ave. 2
science park
hsinchu city, 30075
TW   30075
MDR Report Key24197199
Report Number9681642-2026-00003
Device Sequence Number9979985
Product Code LXH
UDI-Device Identifier04719886892130
UDI-Public(01)04719886892130(10)22G860
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9404-2903
Device Lot Number22G860
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/13/2026
Initial Report FDA Received Date01/28/2026
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-