| Catalog Number |
5833690 |
| Medical Device Problem Codes |
Fracture (1260); Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, a patient underwent an ultrasound-guided dialysis catheter placement procedure via the right internal jugular vein, using a hemostar hemodialysis catheter.During the procedure, it was observed that the aspirated blood revealed the presence of gas.Subsequent inspection of the puncture needle revealed a crack in the handle, causing air leakage and rendering the needle unusable.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, a patient underwent an ultrasound-guided dialysis catheter placement procedure via the right internal jugular vein, using a hemostar hemodialysis catheter.During the procedure, it was observed that the aspirated blood revealed the presence of gas.Subsequent inspection of the puncture needle revealed a crack in the handle, causing air leakage and rendering the needle unusable.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported air/gas in device and needle handle fracture issues as no objective evidence was provided for review.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.D4 (unique identifier (udi) #), g3, h6 (component, method).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|