A facility reported a microsensor (id: 826631) was implanted.The icp express, 220 volt (id: 826635) showed 175 mmhg.However, all post-examination indicators of the patient were normal.Then the physician connected the microsensor to another icp monitor; it still showed 170 mmhg, which was inconsistent with the patient¿s status.Therefore, the sensor was removed on (b)(6) 2026, and monitoring was stopped.
|
|
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h11.The microsensor (id 826631) was returned for evaluation: device history record (dhr) - the product code 826631 with lot 7525928 sn (b)(6), conformed to the specifications when released to stock.Failure analysis - the issue of the complaint was not confirmed: received with sutures, catheter kinked 15.5 cm and 28.4 cm from the distal tip.Debris encapsulated within the rtv.Icp express = 493, microsensor detected and zeroed without issue.The device passed electronic, noise and signal drift tests.Root cause analysis - the exact issue reported by customer could not be confirmed as the device worked correctly.However, the possible root cause of the defect reported by the customer could be due to incorrect set-up of device.
|