Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Correction: d4 (lot #, expiration date and primary udi number) & h4.Investigation ¿ evaluation.It was reported a cook ncircle delta wire tipless stone extractor was defective during an unspecified procedure.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.Reviews of documentation including the complaint history, device history record (dhr), quality control (qc) procedures, and instructions for use (ifu), as well as a visual examination and functional testing of the returned device, were conducted during the investigation.The complaint device was returned in an open package with label.Functional testing was done and the handle would not move at all.The handle was disassembled and visual examination noted that there appears to be adhesive where the cannula enters the sheath at the support sheath flare.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) found no related non-conformances that could have contributed to the reported failure mode.A complaint history database search showed no other related complaint associated with the failure mode for the complaint device lot.Based on the available information, cook has concluded that the device was manufactured to specification and that there is no evidence suggesting nonconforming product exists either in house or in the field.Cook also reviewed product labeling.The ifu packaged with the device did not provide any information related to the reported issue.Based on the available information, inspection of returned device, and the results of the investigation, cook has concluded that the definitive cause of the issue could not be determined.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|