• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON & CO. (SPARKS) BD BACTEC¿ FX40 INSTRUMENT; SYSTEM, BLOOD CULTURING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON & CO. (SPARKS) BD BACTEC¿ FX40 INSTRUMENT; SYSTEM, BLOOD CULTURING Back to Search Results
Catalog Number 442296
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/05/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported while using bd bactec¿ fx40 instrument, there were two false negative results.No adverse impact was reported regarding the patient or user.
 
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Additional Manufacturer Narrative
The following fields have been updated with corrected and/or additional information.B5.Describe event or problem: it was reported while using bd bactec¿ fx40 instrument, there was one false negative result.No adverse impact was reported regarding the patient or user.Investigation summary: a complaint alleges the bactec fx40 instrument had ¿discrepancy between blood culture bottle and tsa agar results¿.Remote services contacted customer to obtain additional information, customer noted that the agar results showed the presence of microbial contamination, however, remote services tested for retention samples and the results were not duplicated.The contamination was not confirmed by the bactec bottles.After 14 days of incubation, the results remain negative.The instrument is working properly.Complaint is unconfirmed based on the information provided by remote services.The root cause is customer workflow.No physical samples were received as part of this complaint; however, pictures and reports were provided and reviewed.Dhr review is not required for this complaint as this complaint does not allege an early life failure or failure at installation and the configuration has changed since release from manufacturing due to service repairs/pms.Service history review for this instrument was completed and revealed no previous complaints related to this issue.Quality will continue to monitor for trends associated with the failure mode described in this complaint.Review of risk management files confirms there are no new or modified risks associated with this failure mode.
 
Event or Problem Description
It was reported while using bd bactec¿ fx40 instrument, there was one false negative result.No adverse impact was reported regarding the patient or user.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BACTEC¿ FX40 INSTRUMENT
Common Device Name
SYSTEM, BLOOD CULTURING
Manufacturer (Section D)
BECTON DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
matt j. connor
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key24199983
Report Number1119779-2026-00029
Device Sequence Number17516666
Product Code MDB
UDI-Device Identifier00382904422967
UDI-Public(01)00382904422967
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K915796
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number442296
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2026
Supplement Date Received by Manufacturer02/25/2026
Initial Report FDA Received Date01/28/2026
Supplement Report FDA Received Date03/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/10/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-