| Model Number |
SC-2318-50 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/08/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Block b3: approximated based on the date the manufacturer became aware of the event.Block d2b: lgw, qrb.Additional suspect medical device components involved in the event: product family: scs-linear leads: upn: m365sc2318500, model: sc-2318-50, serial: (b)(6), batch: 5004689, udi: (b)(6).Product family: scs-linear leads: upn: m365sc43180, model: sc-4318, batch: m365sc43180, udi: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's spinal cord stimulator (scs) leads had migrated which was confirmed through x-ray.The patient underwent a lead replacement procedure and was doing well postoperatively.The explanted leads were retained by the medical facility and will not be returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's spinal cord stimulator (scs) leads had migrated which was confirmed through x-ray.The patient underwent a lead replacement procedure and was doing well postoperatively.The explanted leads were retained by the medical facility and will not be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: all explanted devices were not returned for analysis; therefore, a technical analysis could not be performed.Device history record: it was confirmed that all devices met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review: review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion: all explanted devices were not returned for analysis and the complaint as reported could not be confirmed.Record review revealed no additional information related to the complaint.Therefore, this investigation is unable to determine a probable cause for the complaint and the conclusion code known inherent risk of device will be used.
|
| |
|