|
Continuation of d10: 5071-53 lead implanted (b)(6) 2014, dtba1d4 icd implanted (b)(6) 2018 this information was received from the mechanical circulatory support product surveillance registry study.Investigation completion.Product event summary: (b)(6) was not returned for evaluation.A review of the manufacturing documentation was not performed as the reported event is not related to a manufacturing or servicing issue.The reported suction event could not be confirmed via review of the controller log files since log files were not available for analysis.Based on the available information, the device may have caused or contributed to the reported event.Based on the risk documentation, possible causes of the reported suction event may be attributed to multiple factors including but not limited to thrombus at the inflow cannula, poor vad filling, and/or inappropriate pump rotational speed.Per the instructions for use, worsening heart failure and right ventricular failure are known potential complications associated with the implantation of a vad.Possible clinical factors that may have contributed to this event include the patient¿s pre-existing history and related comorbidities, progression of their underlying disease, issues related to the therapeutic use of anticoagulant and antiplatelet medications and the patient's complex post-operative course.There are possible patient, pharmacological and procedural factors that may have contributed to this event.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|