STRYKER-ENDOSCOPY LAKELAND 5300; ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R
|
Back to Search Results |
|
| Catalog Number |
LF1837 |
| Medical Device Problem Code |
Failure to Seal (4070)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
12/01/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a lap assist vaginal hysterectomy with bilateral salpingo oophorectomy, the surgeon said the ligasure was not sealing as it typically did through the case.End tone and energy delivery was noted; no re-grasp alert occurred.The cleaning of the jaws of the ligasure handpiece was performed frequently, as always.Initially, several good seals were completed successfully, however, by the time the procedure reached the uterine and cervix, the previous seals began to come apart and started bleeding so much that the procedure had to convert to open.It was also reported that a a different device was used and it worked fine according to the surgeon.As reported, the tissue bundle was not thick, as the uterine tissues had been skeletonized.Patient experienced a blood loss of 900 milliliters and will be given a blood transfusion once on the floor of the hospital.Patient hospitalization was extended of one week.Patient had a small bowel obstruction during this time too.The patient is currently healing, though the recovery process has been lengthy.The generator used in the event worked okay with other procedures/ligasures devices.There was no other patient injury or medical intervention reported and the complainant is not aware of any extended procedure time.These are commonly used devices that are readily available.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned to stryker sustainability solutions for evaluation.The submitted medwatch form did not specify the availability of the device for return.As the device was not returned for evaluation, inspection was unable to be performed.The customer did not report lot # or serial # information.Therefore, the device history record (dhr) was unable to be verified.Stryker procedure(s) support(s) that the device was unlikely to have been released from stryker with the reported failure mode.The reported event could be attributed to: - spacer pad damage - ancillary equipment failure - device activated in contact with pooling fluid.- device used on vessels thicker than 7 mm.- grasping on too much tissue or inappropriate tissue types or on staples.- trigger button (activation button) not engaged throughout the entire seal cycle.The instructions for use (ifu) state: - do not place the vessel and/or tissue in the jaw hinge.Place the vessel and/or tissue in the center of the jaws.- do not use this instrument on vessels larger than 7 mm in diameter.- if the instrument shaft is visibly bent, discard and replace the instrument.A bent shaft may prevent the instrument from sealing or cutting properly.- eliminate tension on the tissue when sealing and cutting to ensure proper function.- use caution when grasping, manipulating, sealing, and dividing large tissue bundles.- do not attempt to seal or cut over clips or staples as incomplete seals will be formed.Contact between an active electrode and any metal objects may result in alternate site burns or incomplete seals.- do not activate the ligasure system until the lever has been latched.Activating the system before latching may result in improper sealing and may increase thermal spread to tissue outside of the surgical site.- prior to cutting the seal, inspect the vessel or tissue to ensure proper sealing.- keep the instrument jaws clean.Build-up of eschar may reduce the seal and/or cutting effectiveness.Wipe jaw surfaces and edges with a wet gauze pad as needed.- if the generator provides multiple power settings, use the lowest power needed to achieve the intended effect.- do not overfill the jaws of the instrument with tissue, as this may reduce device performance.- use caution during surgical procedures in which patients exhibit certain types of vascular pathology (atherosclerosis, aneurysmal vessels, etc.).For best results, apply the seal to unaffected vasculature.- energy-based devices, such as es pencils or ultrasonic scalpels that are associated with thermal spread, should not be used to transect seals.The reported event will continue to be monitored through post-market surveillance.Should the device become available for return, the investigation will be reopened.
|
| |
|
Search Alerts/Recalls
|
|
|