| Model Number |
TRIEEAXL25MT |
| Medical Device Problem Codes |
Device Difficult to Setup or Prepare (1487); Noise, Audible (3273)
|
| Health Effect - Clinical Codes |
Failure to Anastomose (1028); Tissue Breakdown (2681)
|
| Date of Event |
01/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic procedure roux en y, when creating gastric pouch, the white tip had fallen out of anvil before being inserted, and the surgeon could not get it to stay back in the anvil so proceeded without using the white accessory tip.Then, while firing, after the surgeon completed all the correct steps for firing, when squeezed the handle, the user did not hear an audible click nor could really feel anything, but the handle was completely squeezed.There was also a torn spot near the staple line that the surgeon then stitched up.To resolve the issue, another circular stapler was used.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that when creating gastric pouch, the white tip had fallen out of anvil before being inserted, and the surgeon could not get it to stay back in the anvil so proceeded without using the white accessory tip.Then, while firing, after the surgeon completed all the correct steps for firing, when squeezed the handle, the user did not hear an audible click nor could really feel anything, but the handle was completely squeezed.There was also a torn spot near the staple line that the surgeon then stitched up.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|