Model Number DRT225
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Disturbances (2140); Halo (2227)
Date of Event 09/16/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Section d6b: if explanted, give date: not applicable, as the lens remains implanted.Device evaluation: the device was not returned to the manufacturing site, as it remains impanted; therefore, product testing could not be performed, and the reported complaint could not be verified.Manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.A search of complaints related to this production order (po) was performed.The search revealed that no similar complaints for this po number have been received.Conclusion: based on the complaint investigation results, the product was released within specifications.The complaint issue reported could not be confirmed and no product deficiency or product malfunction could be identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported by the patient that following the implantation of a preloaded multifocal toric intraocular lens (iol) in the right and left eye, the patient experienced persistent halos around lights in both eyes.The physician had indicated that the halos would resolve within a month, then six months, and eventually over time; however, the issue has persisted for 15 months without improvement.The patient described seeing one or two circles of light around every light source, regardless of brightness or time of day, which affects daily activities such as night driving and viewing lighted decorations during holidays.The halos interfere with visibility of turn signals and seeing the backs of vehicles ahead while driving at night and are disruptive during daytime activities.The patient also reported suboptimal near and arm's length vision, occasionally requiring readers for tasks such as laptop work.The lenses remain implanted, and no medical or surgical intervention has been performed or planned.No further information was provided.This report is capturing the event for the right eye.A separate report will capture the event for the left eye.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ODYSSEY TORIC II IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key24206137
Report Number3012236936-2026-000028
Device Sequence Number5534590
Product Code MFK
UDI-Device Identifier05050474811799
UDI-Public(01)05050474811799(17)270716
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberDRT225
Device Catalogue NumberDRT2250125
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/07/2026
Initial Report FDA Received Date01/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/16/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age73 YR
Patient SexMale
Patient Weight88 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-