| Model Number |
CL-803 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: cl-803; cl-803 polysorb* 1 vio 75cm gs24 x36; (lot number: a5g1541y) cl-803; cl-803 polysorb* 1 vio 75cm gs24 x36; (lot number: a5g1541y).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during the total abdominal hysterectomy, three needles detached from the suture while suturing.A suture from other manufacturer was used to complete the case.There was no patient injury.
|
| |
|
Search Alerts/Recalls
|