| Model Number |
470001-09 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/22/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) has not received the da vinci product with an alleged issue to perform failure analysis.A follow-up mdr will be submitted if the product is returned and evaluated and/or if additional information is received.
|
| |
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, the 8mm potts scissors instrument tips are not closing all the way.The procedure was completed with no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical inc.(isi) received the 8mm potts scissors instrument for failure analysis investigation.The instrument was analyzed and found to have one blade bent at the tip.The blades were not able to fully close as a result of the bending.Cut performance is negatively impacted due to the bending.The complaint was confirmed by failure analysis.The probable root cause of bent blades is attributed to damage during use, as moderate misalignment of tips may result from attempting to grasp either hard tissue/objects or inadvertent collisions with other instruments.Applying excessive force on the instrument tips during handling, insertion, usage (overloading), or removal can also cause bending.
|
| |
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|
Search Alerts/Recalls
|