• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 REPAIR WHITE ADAPTER LEG; PORT AND CATHETER ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 REPAIR WHITE ADAPTER LEG; PORT AND CATHETER ACCESSORIES Back to Search Results
Catalog Number 0601680
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
On (b)(6) 2026, a broviac catheter was repaired using the hickman/leonard/broviac repair kit.It was reported that during the repair, it was noted that the small metal piece, which allegedly connects the repair to the line, was not stable or properly glued in place.The component was movable, raising concern from the attending nurse that it could potentially flush off into the line or patient if the line were flushed with excessive force.There was no reported patient injury.
 
Event or Problem Description
On (b)(6) 2026, a broviac catheter was repaired using the hickman/leonard/broviac repair kit.During the repair, it was noted that the small metal piece, which allegedly connects the repair component to the line, was not stable or properly secured in place.The component was movable, raising concern from the attending nurse that it could potentially flush off into the line or the patient if the line were flushed with excessive force.There was no reported patient injury.
 
Additional Manufacturer Narrative
H11: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.The investigation is inconclusive for the reported loose or intermittent connection issue as no objective evidence was provided for the review.The definitive root cause could not be determined based upon available information.Labelling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.G3, h6 (component, method).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REPAIR WHITE ADAPTER LEG
Common Device Name
PORT AND CATHETER ACCESSORIES
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key24207258
Report Number3006260740-2026-00286
Device Sequence Number14810369
Product Code LJS
UDI-Device Identifier00801741036415
UDI-Public(01)00801741036415(17)280131(10)REKW0354
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K830406
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0601680
Device Lot NumberREKW0354
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/06/2026
Supplement Date Received by Manufacturer03/04/2026
Initial Report FDA Received Date01/29/2026
Supplement Report FDA Received Date03/10/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1