MEDTRONIC MEXICO S. DE R.L. DE CV CUSTOM TUBING PACK; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO
|
Back to Search Results |
|
| Model Number |
BB8V35R8 |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received information that prior to use of a custom packs, it was reported that the outer packaging was damaged.The devices were not used.There was no patient involvement, so no adverse effect occurred.Medtronic received additional information that there were no missing or wrong devices or components in the package.The customer was unsure if the sterile seal was intact.The device was partially sealed/located in the seal.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: visual inspection of the unopened device showed damage to the outer package.Additional visual inspection showed evidence of damage to the tyvek lid.Reason for return was confirmed.This regulatory report is being submitted as part of a retrospective review and remediation as part of a capa action.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|