• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN EEA; STAPLER, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN EEA; STAPLER, SURGICAL Back to Search Results
Model Number TRIEEAXL25MT
Medical Device Problem Code Incomplete or Inadequate Connection (4037)
Health Effect - Clinical Codes Failure to Anastomose (1028); Abdominal Pain (1685); Laceration(s) (1946)
Date of Event 01/07/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10 - concomitant product: eeaorvil25a, eeaorvil25a eea orvil 25mm automaticdv, (lot# n4j1907y).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, intra-operatively of a total gastrectomy, during the anastomosis of the esophageal stump and the elevated jejunum, the circular stapler handle and the oral anvil could not be attached which resulted in the anastomosis being unable to be performed.The procedure was changed from laparoscopy to laparotomy and thoracotomy.The surgical time was extended for three hours.It was reported that the patient had acute abdominal pain post-operatively.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the circular stapler handle and the oral anvil could not be attached which resulted in the anastomosis being unable to be performed.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EEA
Common Device Name
STAPLER, SURGICAL
Manufacturer (Section D)
COVIDIEN
sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
COVIDIEN
sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24209853
Report Number2647580-2026-00216
Device Sequence Number5430816
Product Code GAG
UDI-Device Identifier10884521593107
UDI-Public(01)10884521593107(17)280731(10)P3H1251
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K172361
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTRIEEAXL25MT
Device Catalogue NumberTRIEEAXL25MT
Device Lot NumberP3H1251
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/07/2026
Supplement Date Received by Manufacturer03/25/2026
Initial Report FDA Received Date01/29/2026
Supplement Report FDA Received Date04/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/31/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age80 YR
Patient SexMale
-
-