• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR); MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR); MITRAL VALVE REPAIR DEVICES Back to Search Results
Model Number CDS0706-XT
Medical Device Problem Codes Unstable (1667); Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been received.However, investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported on 06 january 2026 that the patient presented with grade 4 degenerative mitral regurgitation (mr) and prolapsed leaflet for a mitraclip procedure.During the procedure, the clip was unable to pass establish final arm angle test (efaa).The arm positioner felt unstable.The clip was removed from the anatomy.The clip was replaced with another device to complete the procedure.The mr was reduced to grade 1-2 post procedure.There was no clinically significant delay or adverse patient effects.
 
Additional Manufacturer Narrative
All available information was investigated, and the reported unstable arm positioner and unintended movement of clip open during efaa were not confirmed via returned device analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.The investigation was unable to determine a cause for the reported unstable arm positioner and unintended movement of clip open during efaa.There is no indication of a product issue with respect to manufacture, design, or labeling.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR)
Common Device Name
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL NORTHPOINT REG 3020950818
1820 bastian court
westfield IN 46074
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key24210848
Report Number2135147-2026-00473
Device Sequence Number17526382
Product Code NKM
UDI-Device Identifier05415067037404
UDI-Public(01)05415067037404(17)260429(10)50430A1078
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P100009
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/29/2026
Device Model NumberCDS0706-XT
Device Catalogue NumberCDS0706-XT
Device Lot Number50430A1078
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/06/2026
Supplement Date Received by Manufacturer03/02/2026
Initial Report FDA Received Date01/29/2026
Supplement Report FDA Received Date03/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/30/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
STEERABLE GUIDE CATHETER.
Patient Age72 YR
Patient SexMale
-
-