| Model Number |
CDS0706-NTW |
| Medical Device Problem Code |
Incomplete Coaptation (2507)
|
| Health Effect - Clinical Code |
Mitral Valve Insufficiency/ Regurgitation (4451)
|
| Date of Event |
12/23/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device remains implanted in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2025, a patient presented with grade 4 degenerative mitral regurgitation (mr) and prolapsed posterior leaflet for a mitraclip procedure.During the procedure, two mitraclips were implanted.All steps were performed as per ifu.The final mr was grade <1.There were no adverse patient effects or clinically significant delays reported.On (b)(6) 2025, it was determined that a single leaflet device attachment has occurred for both the clips and clips were no longer attached to the posterior leaflet.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported incomplete coaptation.Additionally, a review of the complaint history identified no similar complaints from the lot.Based on available information, a cause for the reported incomplete coaptation- single leaflet device attachment (slda) appears to be related to patient morphology/pathology.The reported recurrent mitral valve insufficiency/ regurgitation appears to be a cascading effect of the reported slda.Mr is listed in the instructions for use as a known possible complication associated with the mitraclip procedures.The reported serious injury/ illness/ impairment was a result of case-specific circumstances as no intervention was reported.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Search Alerts/Recalls
|