Model Number 350-013-000
Medical Device Problem Code Material Opacification (1426)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/14/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information: d9, g3, h3 and h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted no physical damage on the received videolaryngoscope blade.The blade was connected to a test videolaryngoscope and powered on.The display functioned properly, with no issues in fit or glare from the blade, and no visual artifacts were observed through the videolaryngoscope camera.The videolaryngoscope was then installed in a cap for a beaker filled with water, where in-air temperature was heated to approximately 35°c as measured for anti-fog testing.Fogging of the viewing prism was observed while the videolaryngoscope was installed for approximately 30 seconds, viewed through the videolaryngoscope's monitor, until the anti-fog coating began operating properly and gradually clearing the fog from the viewing prism.Immediate visual investigation of the blade of the videolaryngoscope after removal from the beaker cap found there to be no condensation on the blade prism.All later installations into the beaker cap to verify resulted in no further fogging of the videolaryngoscope's camera.It was reported that moisture generation was observed with the videolaryngoscope blade prior to use, at a hospital.The reported issue was confirmed.The most likely cause was traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that moisture generation was observed with the videolaryngoscope blade prior to use, at a hospital.There was no pat ient involved.
 
Additional Manufacturer Narrative
D10 concomitant product: 350-013-000, mcgrath mac 4 blade 350-013-000 50 ct, lot# 24k0098jzx.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCGRATH MAC
Common Device Name
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX  32575
Manufacturer (Section G)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX   32575
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24215084
Report Number9681384-2026-00166
Device Sequence Number21141810
Product Code CCW
UDI-Device Identifier10884521824324
UDI-Public(01)10884521824324(17)271101(10)24K0098JZX
Combination Product (Y/N)N
Initial Reporter CountryKS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number350-013-000
Device Catalogue Number350-013-000
Device Lot Number24K0098JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/14/2026
Supplement Date Received by Manufacturer04/03/2026
Initial Report FDA Received Date01/29/2026
Supplement Report FDA Received Date04/16/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NOTES ON H11
Patient SexUnknown
-
-