| Model Number |
102956 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that an s3 alarm occurred during a patient transport within a hospital.The customer stated that while transporting a centrimag venovenous extracorporeal membrane oxygenation (vv-ecmo) patient, the centrimag console was plugged into a battery backup/power strip device.The s3 alarm did not resolve when they plugged the battery backup device into a hospital emergency outlet.The customer was advised to discontinue the use of the power strip and to switch to another centrimag console and motor that was plugged directly into a red hospital ac power outlet.The customer stated that they did not want to switch to another device at this time as the patient was considered a do not resuscitate (dnr) and that somehow switching to another device would be a problem.Several hour later the customer switched to another centrimag console yet continued to use the battery backup/power strip device and the 2nd centrimag system also displayed an s3 alarm.
|
| |
|
Additional Manufacturer Narrative
|
|
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.
|
| |
|