|
Medtronic received a report that upon removing the navien catheter from its packaging, a kink was discovered in the distal end.The product was replaced with a new navien catheter to complete the procedure.There was no patient involvement.The patient was undergoing coil embolization surgery for treatment of an right posterior communicating artery aneurysm.It was noted the patient's vessel tortuosity was normal.Femoral artery access vessel diameter was 8 mm.The reported device and any accessory devices were prepared, and the catheter was flushed, as indicated in the instructions for use (ifu).Additional information was received stating that the cause of the kink was not determined.There was no damage or evidence of tampering to device packaging.The device packaging was not damaged on delivery.
|
|
Product analysis as found condition: the navien was returned within a shipping box and a plastic bio-pouch.Visual inspection/damage location details: no damage was found to the navien catheter hub.The navien catheter body was found to be broken at 104.1cm from proximal end of hub.The naviend distal tip was found to be in good condition.Testing/analysis: the navien catheter total length was measured to be 122.5cm and the usable length was measured to be 115.7cm, which is within specification.Conclusion: based on the device analysis and reported information, the customer¿s ¿catheter separation/break¿ and ¿prod damaged/deformed out of pkg¿ were confirmed as the returned navien was broken along the catheter body.However, ¿catheter kink/damage¿ could not be confirmed.Upon inspection, a broken section was observed near the distal end of the catheter.The damage of the catheter may have occurred due to an impact from unknown sources prior to being opened, damage during storage, use-related, manufacturing related or hand delivery.However, the root cause of the damaged catheter could not be determined.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|