| Catalog Number |
8065752182 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that there was an abnormal sound during cataract surgery with intraocular lens implantation.The surgery was completed on the same day without product replacement.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in h.6.And h.11.Product evaluation was not able to be performed since the reported product was not received at the investigation site for evaluation.The investigation conducted a non-conformance review of the reported lot number.No deviations that could have contributed to this event were identified and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information, the investigation was unable to conclusively identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.No further investigative or manufacturing actions are warranted at this time due to the product not being returned to the manufacturing site for testing.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|