• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. MCRYL VIO 27IN 2-0 S/A UR-6; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. MCRYL VIO 27IN 2-0 S/A UR-6; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number Y605H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/10/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2025 and suture was used.When the needle was placed in the troluca, it was found that the problem of pulling off suture needle happened.The needle was found immediately and its integrity was confirmed.There were no patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 1/30/2026.H6.Component code: g07002 - no device problem found.H6.Component code: c22 - photo analysis.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, the following: did the reported issue contribute to any patient adverse consequences? when was the needle detached/pulled off from the suture (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify.Received information as following from sales rep via phone today.Please refer to the event description, other information requested is unknown.Photo analysis: this is an analysis for a photo submitted to ethicon for evaluation.During the visual analysis, the following was observed: the picture shows a needle without suture inside a plastic bag.A clear analysis of the end point of failure or the needle hole could not be performed due to the position of the needle.Based on the photo review, the event reported is confirmed, however no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the device was not returned our evaluation is limited.As part of ethicon quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCRYL VIO 27IN 2-0 S/A UR-6
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key24220744
Report Number2210968-2026-01100
Device Sequence Number8085426
Product Code GAN
UDI-Device Identifier10705031059481
UDI-Public(01)10705031059481(17)290131(10)UBMEJK
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K960653
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberY605H
Device Lot NumberUBMEJK
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/06/2026
Initial Report FDA Received Date01/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-