| Model Number |
Z021135 |
| Medical Device Problem Codes |
Excess Flow or Over-Infusion (1311); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/19/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The following has been reported by customer: pump infused medication at fast rate.Customer emailed in the request form reporting: pump infused medication at rate faster than it was programmed for.Reporting as a conservative measure.Adverse effects were not reported.Additional information is needed to complete the investigation.
|
| |
|
Event or Problem Description
|
|
Device history record was reviewed.No event linked with the reported issue was observed.Device log was reviewed, no error or issue linked to the reported event was found.The reported device was received in lake zurich and its investigation was performed by our local technical team.According to provided information, nothing was noticed during both external and internal check (except the externals, internals and inseams found dirty).Several functional tests related to the reported event were carried out.All performed successfully and values were found compliant with ifu specifications compared to settings.The reported event could not be reproduced.Based on available information, this complaint is considered as not valid with no issue.To our knowledge this complaint is not being related to patient safety.This complaint is considered as: not valid.The trend is: n/a.
|
| |
|
Search Alerts/Recalls
|