• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WATCHMAN FLX? PRO; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WATCHMAN FLX? PRO; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL Back to Search Results
Model Number M635WU60400
Medical Device Problem Code Failure to Seal (4070)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Reported via clinical study: it was reported device leaks occurred.A left atrial appendage (laa) closure procedure was performed and a 40mm watchman flx pro closure device was implanted.On a routine follow up, 12 months post index procedure, computed tomography (ct) imaging revealed 2 leaks around the closure device, with the first leak measuring 8.2mm and the second leak measuring 5.1mm.There were no corrective actions or diagnostic tests associated with the finding.
 
Event or Problem Description
Reported via clinical study: it was reported device leaks occurred.A left atrial appendage (laa) closure procedure was performed and a 40mm watchman flx pro closure device was implanted.On a routine follow up, 12 months post index procedure, computed tomography (ct) imaging revealed 2 leaks around the closure device, with the first leak measuring 8.2mm and the second leak measuring 5.1mm.There were no corrective actions or diagnostic tests associated with the finding.
 
Additional Manufacturer Narrative
With all the available information, boston scientific (bsc) concludes: device technical analysis: the watchman flx? pro remains implanted; therefore, returned device analysis could not be completed.Device history record review: a review was completed of the device history records (dhr) for the watchman flx? pro, part # m635wu60400 batch # 0032255230.The device was processed through all normal operational conditions and passed all acceptance activities.The dhr review did not identify any deviations or non-conformances that could have contributed to the event.Labeling review: there was no evidence to suggest that the device was used in a manner inconsistent with the labeling.Watchman flx? pro instructions for use (ifu) lists potential complications, risks and adverse events that may be associated with the use of this device which include implant did not seal.Risk review: a risk review was completed and confirmed that the event of implant did not seal was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN FLX? PRO
Common Device Name
SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key24221697
Report Number2124215-2026-05484
Device Sequence Number9366410
Product Code NGV
UDI-Device Identifier00191506004637
UDI-Public(01)00191506004637(17)260820(10)32255230
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
P130013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/20/2026
Device Model NumberM635WU60400
Device Catalogue NumberM635WU60400
Device Lot Number0032255230
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/27/2026
Supplement Date Received by Manufacturer03/19/2026
Initial Report FDA Received Date01/30/2026
Supplement Report FDA Received Date03/26/2026
Date Device Manufactured08/21/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age80 YR
Patient SexFemale
-
-