| Model Number |
M635WU60400 |
| Medical Device Problem Code |
Failure to Seal (4070)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/15/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Reported via clinical study: it was reported device leaks occurred.A left atrial appendage (laa) closure procedure was performed and a 40mm watchman flx pro closure device was implanted.On a routine follow up, 12 months post index procedure, computed tomography (ct) imaging revealed 2 leaks around the closure device, with the first leak measuring 8.2mm and the second leak measuring 5.1mm.There were no corrective actions or diagnostic tests associated with the finding.
|
| |
|
Event or Problem Description
|
|
Reported via clinical study: it was reported device leaks occurred.A left atrial appendage (laa) closure procedure was performed and a 40mm watchman flx pro closure device was implanted.On a routine follow up, 12 months post index procedure, computed tomography (ct) imaging revealed 2 leaks around the closure device, with the first leak measuring 8.2mm and the second leak measuring 5.1mm.There were no corrective actions or diagnostic tests associated with the finding.
|
| |
|
Additional Manufacturer Narrative
|
|
With all the available information, boston scientific (bsc) concludes: device technical analysis: the watchman flx? pro remains implanted; therefore, returned device analysis could not be completed.Device history record review: a review was completed of the device history records (dhr) for the watchman flx? pro, part # m635wu60400 batch # 0032255230.The device was processed through all normal operational conditions and passed all acceptance activities.The dhr review did not identify any deviations or non-conformances that could have contributed to the event.Labeling review: there was no evidence to suggest that the device was used in a manner inconsistent with the labeling.Watchman flx? pro instructions for use (ifu) lists potential complications, risks and adverse events that may be associated with the use of this device which include implant did not seal.Risk review: a risk review was completed and confirmed that the event of implant did not seal was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
|
| |
|
Search Alerts/Recalls
|