RESPIRONICS, INC. BIPAP A40; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
| Model Number |
INX3100S19 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The bipap a40 pro (inx3100s19) is substantially similar to the bipap a40 (1111169) and will be reported in the united states under bipap a40, 501k number: k121623.
|
| |
|
Event or Problem Description
|
|
The bipap a40 pro ventilator was returned to the third-party service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There was no allegation of serious or permanent harm or injury.The device was returned to the third-party service center for investigation.During evaluation, evidence of (foam degradation inside the assembly, deteriorated foam) was present.There were no repairs and the device was scrapped.
|
| |
|