The customer reported a falsely elevated tacrolimus (tac) result was obtained on 1 patient sample on a dimension exl 200 integrated chemistry system.The result was not reported to the physician(s).The sample was repeated on the same dimension exl 200 integrated chemistry system.The repeat result was lower than the erroneous result.The repeat result was reported, as the correct result, to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the falsely elevated tac results.
|
|
Siemens filed the initial mdr 2517506-2026-00017 on 30-jan-2026.Additional information (31-mar-2026): siemens further evaluated the event by reviewing the available information in the complaint.Evaluation included review of the analyzer performance, quality control data, calibration status, and raw reaction absorbance measurements.Quality control testing performed after the event recovered within acceptable laboratory ranges, confirming stable reagent and calibration performance.In addition, review of the tacrolimus (tac) assay reaction filter data.577 nm and 700 nm, demonstrated stable background and analytical absorbance progression across reaction points, indicating normal photometric performance and consistent reaction kinetics without evidence of optical instability, abnormal reaction curves, or cuvette-related issues.System checks and analyzer operational parameters were also verified to be within expected specifications during the reported testing period.Based on the available investigation findings, there is no evidence of analyzer malfunction.Considering that tac utilizes ethylenediaminetetraacetic acid (edta) whole blood and relies on adequate red blood cell lysis and homogeneous sample mixing prior to aspiration, the observed repeat variability is most consistent with sample-specific factors inherent to whole-blood testing, such as partial sample settling, variability in red blood cell distribution, or incomplete sample homogenization prior to aspiration.The complaint was resolved with routine troubleshooting actions.The customer is operational.The investigation findings and investigation conclusions in section h6 were updated to reflect the additional information.No further evaluation of this analyzer is required mdr's 2517506-2026-00006, 2517506-2026-00018, 2517506-2026-00019, 2517506-2026-00020, 2517506-2026-00021, 2517506-2026-00022, 2517506-2026-00023, 2517506-2026-00024, 2517506-2026-00025, 2517506-2026-00026, 2517506-2026-00027, 2517506-2026-00028, 2517506-2026-00029, 2517506-2026-00030, 2517506-2026-00031, and 2517506-2026-00032 were filed for the same event.
|