| Model Number |
KD-V611M-07251S |
| Medical Device Problem Codes |
Break (1069); Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported the single use preloaded sphincterotome had a broken wire.The issue occurred during a therapeutic endoscopic retrograde cholangiopancreatography and was completed with an olympus back-up device.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated fields: d9, h2, h3, h6.The device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: coating on the cutting wire was peeled and dislodged.Based on the results of the investigation a root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|