| Catalog Number |
0375544000 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the shaver attachment fractured at the cutting blade, rendering it unusable.A spare device was available to successfully complete the medical procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported the customer disposed of the device.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: shaver attachment fractured at the cutting blade.Probable root cause: inadequate window profile.Inadequate welding process (i.E.Plasma, inductive, gluing).Improper material strength.Inadequate size selected for the application.Material brittleness.Excessive force applied by the user.Incorrect rfid tag".The reported failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the shaver attachment fractured at the cutting blade, rendering it unusable.A spare device was available to successfully complete the medical procedure.
|
| |
|