MAUDE Adverse Event Report: SYNTHES GMBH RETENTION PIN FOR SCREWDRIVER SHORT / XL25

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH RETENTION PIN FOR SCREWDRIVER SHORT / XL25 Back to Search Results
Catalog Number 03.045.004
Medical Device Problem Codes Break (1069); Entrapment of Device (1212)
Health Effect - Clinical Codes Foreign Body In Patient (2687); Insufficient Information (4580)
Date of Event 01/20/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2026, the patient underwent orif surgery with tna for tibial shaft fractures.During insertion of a low-profile screw into the fourth ml hole from the distal end, the retention pin broke off when the screw head was partially embedded.The broken tip remained in the screw head, but the surgeon closed the wound as is because there was no problem with fixation.The surgery was completed successfully within thirty minutes of delay.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d9 the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: corrected: g1.H3, h6: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device to be broken from the tip.The broken fragment was not returned.No postoperative x-rays were provided to confirm the embedded device, therefore this allegation can't be confirmed.No other issues were observed.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.A dimensional inspection was unable to be performed due to the post-manufacturing damage.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the retention pin short xl25 would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: part number: 03.045.004.Lot number: 5544p66.Manufacturing site: haegendorf.Release to warehouse date: 26-jun-2023.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances related to the malfunction were identified.Finished device lot number is part of capa jbl-capa-000388, however, this capa is not related to the pi, but to the application of secondary passivation process using citric acid instead of nitric acid which is approved as use-as-is and documented accordingly in jbl-nr / jnj-nr-0199988.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RETENTION PIN FOR SCREWDRIVER SHORT / XL25
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK HAEGENDORF
im bifang 6
haegendorf CO
SZ  
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key24224551
Report Number8030965-2026-01088
Device Sequence Number9385236
Product Code HXX
UDI-Device Identifier07612334167811
UDI-Public(01)07612334167811(10)5544P66
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.045.004
Device Lot Number5544P66
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/19/2026
Supplement Date Received by Manufacturer02/09/2026
03/01/2026
Initial Report FDA Received Date01/30/2026
Supplement Report FDA Received Date02/10/2026
03/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/23/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-